• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST Back to Search Results
Model Number 100071
Device Problems Incorrect Or Inadequate Test Results (2456); High Test Results (2457)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 11/05/2015
Event Type  malfunction  
Manufacturer Narrative
Investigation pending.
 
Event Description
The caller alleged a variance between inratio inr results and the lab inr results.The results were as follows: (b)(6).
 
Manufacturer Narrative
It is indicated that the product is not returning for evaluation.Therefore, a review of the entire in-house testing history was performed.In-house testing on strip lot 376877a meets release criteria.The product performed as expected.A review of the manufacturing records for lot# 376877a did not uncover any non-conformances.The lot meets release specifications.Root cause could not be determined from the information provided by the customer.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INRATIO PT/INR TEST STRIPS
Type of Device
PROTHROMBIN TIME TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
Manufacturer (Section G)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
Manufacturer Contact
ya-ling king
9975 summers ridge rd.
san diego, CA 92121
8588052084
MDR Report Key5333079
MDR Text Key35062837
Report Number2027969-2015-01033
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 12/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number100071
Device Lot Number376877A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/02/2015
Initial Date FDA Received12/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-