Findings: as of the date of this submission the involved catheter device has not been returned for analysis and confirmation.The dfu provides specific instructions for pre-testing/set-up and handling instructions.The catheter was successfully pre-tested with no functional issues or performance abnormalities detected at that time.The opticath® catheter is a precision optical instrument and must be handled with care.Kinking, excessive manipulation, or grasping with forceps or hemostats can permanently damage the fiberoptics, reducing the transmission of light to and from the blood.
|
Device return: one used 50328-08 8f triox svo2 pa catheter w/ 8.5f contamination shield.Engineering testing and analysis: the returned 50328-08 catheter was tested per the applicable specifications including fiber optic and coupler inspection.The results recorded the catheter did not meet specifications, light did not pass through both the sending or return fiber.Additionally the investigation report described additional fiber optic component damages as evidenced by the break/separation of the fiber optic from the catheter connector tips.The 50328-08 catheter was dissected and the components evaluated.The report documented numerous breaks, in the fiber optic in proximity to the 40, 60, 75 108 cm locations.Breaks were also observed at the catheter lumen junction and the thermistor and fiber optic junction locations.The report notes these types of damages are consistent with excessive tension force, sharp bends or excessive physical impact shocks.The dfu provides specific instructions for pre-testing/set-up and handling instructions.The catheter was successfully pre-tested with no functional issues or performance abnormalities detected at that time.The opticath catheter is a precision optical instrument and must be handled with care.Kinking, excessive manipulation, or grasping with forceps or hemostats can permanently damage the fiberoptics, reducing the transmission of light to and from the blood.(b)(4).Findings: engineering testing and analysis of the returned 50328-08 catheter confirmed the reported product issue.The root cause(s) were determined to be a result of extensive component/fiber optic damages attributable to incorrect usage/technique.
|