CYBERONICS PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Device Problems Use of Incorrect Control/Treatment Settings (1126); Failure of Device to Self-Test (2937)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
A programming history review was performed and a programming anomaly which altered settings was identified.It appears that sometime between (b)(6) 2002 and (b)(6) 2002, a system diagnostic test was performed that changed settings to unintended settings.Settings were programmed on (b)(6) 2002 to without performing a final interrogation.The patient was then seen on (b)(6) 2002 and was interrogated and settings were indicative of a faulted diagnostic test.The settings, output current and off time were then programmed back to intended settings.The physician's tablet in which this occurred and the exact date of occurrence is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5335293
MDR Text Key35171393
Report Number1644487-2015-06867
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Event Location Other
Initial Date Manufacturer Received 12/09/2015
Initial Date FDA Received12/30/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age54 YR
-
-