The device referenced in this report was returned to olympus for evaluation.The evaluation found that there are no sharp edges or burrs on the distal tip.The exact cause of the patient's outcome could not be determined, as there was no device damage that would likely cause or contribute to the reported event.The instruction manual states: "visually inspect the product.Make sure that it has: no corrosion, no dents, no scratches." if significant information is received at a later time, this report will be supplemented.
|