Model Number 10-88-00
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 11/17/2015
Event Type  malfunction  
Manufacturer Narrative
Device has not been returned to sorin group (b)(4).Sorin group (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the s5 mast roller pump displayed an error message during set-up.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the s5 mast roller pump displayed an error message during set-up.There was no patient involvement.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the s5 mast roller pump displayed an error message during set-up.There was no patient involvement.A sorin group service representative was dispatched to the facility to investigate.The service representative evaluated the pump and confirmed the reported failure.The error code was displayed immediately upon turning the pump on.The service representative traced the error to the motor control board.The board was replaced and subsequent functional tests did not identify further issues.The pumpt was returend to service and the motor control board was returned to sorin group (b)(4) for investigation.Visual inspection of the returned device did not identify any abnormalities or defects.The board was connected to the s5 roller pump testbench, however the reported failure could not be reproduced.The supply voltages and pwm signals were tested and found to be within specification.The board was electrically and functionally tested without fault, and thermal testing in a climate chamber did not reproduce the reported issue.An endurance was also performed without fault.All tests performed were failure free.As the issue could not be reproduced, a root cause could not be determined and no corrective actions were identified.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND GMBH
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key5337393
MDR Text Key34739085
Report Number9611109-2015-00595
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K062396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-88-00
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/19/2015
Initial Date FDA Received12/30/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/13/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-