• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Failure to Power Up (1476); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/09/2015
Event Type  malfunction  
Event Description
It was reported that a physician's handheld was not powering on.The battery appeared to have expanded.No patients were affected, because a backup programmer was used.The handheld was received on 12/22/2015.Analysis has not been completed to date.
 
Event Description
Analysis was completed on 01/20/2016 for the handheld and the flashcard.During the analysis it was identified that the handheld would not power on.The cause for the anomaly is associated with a swollen main battery.The swollen battery bent the battery cover causing the battery latch switch to register that the latch was unlocked, thus preventing the handheld from powering on (this condition can also cause the handheld to power off intermittently).No further anomalies were identified.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5337455
MDR Text Key35250357
Report Number1644487-2015-06873
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2015
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 12/09/2015
Initial Date FDA Received12/30/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-