Model Number MERGE HEMO 9.20
Device Problem Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/13/2012
Event Type  malfunction  
Manufacturer Narrative
Upon retrospective review, this issue was determined to be reportable as an mdr.
 
Event Description
Inr values presenting incorrectly in heartsuite hemo - decimal is being dropped resulting in erroneous value.It used an algorithm that looked for a ".0" and replaced it with an empty string.For example: 1.0 would show as 1.If the inr on pre-procedure is 1, it displays on the study report lab section as 10, if the inr on the pre-procedure tab is 1.01 it displays as 11 on the study report lab section.Clinical use may contribute if the physician reading the report does not question the incorrect value and look to the original values.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMO
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge dr.
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge dr.
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge dr.
hartland, WI 53029
2629123570
MDR Report Key5339170
MDR Text Key34886721
Report Number2183926-2015-00021
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial
Report Date 12/31/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMO 9.20
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2012
Initial Date FDA Received12/31/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number2183926-01/04/2016-008-C
Patient Sequence Number1
-
-