Catalog Number 80395
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.No manufacturing problem identified.Rust was found in insert.Inserts are 100% tested in the manufacturing floor fr water flow and spray pattern.Inserts are placed in oven to remove moisture after water test.Customer usage and maintenance is unknown.
 
Event Description
In this event it was reported that a cavitron 30k fsi-sli-10s insert tip overheated; no injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 10S STRAIGHT
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5339501
MDR Text Key34886689
Report Number2424472-2015-00126
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 12/04/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80395
Device Lot Number5296
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/04/2015
Initial Date FDA Received12/31/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-