MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS E-ELECSYS ANTI-CCP; ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS E-ELECSYS ANTI-CCP; ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP) Back to Search Results
Catalog Number 05031656190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received a questionable anti-ccp result for one patient sample from an unknown roche analyzer.The initial result was 120.80 u/ml.The patient was transferred to a different hospital and anti-ccp was tested with stacia meblux test (cleia) on (b)(6) 2015.The result was 0.6 u/ml.The initial result was reported outside the laboratory.The patient was not adversely affected.
 
Manufacturer Narrative
As no sample from the patient was available for further investigation, a specific root cause could not be determined.
 
Manufacturer Narrative
A new sample from the patient was collected and the anti-ccp result on (b)(6) 2016 was 140.8 u/ml.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS E-ELECSYS ANTI-CCP
Type of Device
ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5341809
MDR Text Key35002251
Report Number1823260-2016-00004
Device Sequence Number1
Product Code NHX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K081338
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 03/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05031656190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2015
Initial Date FDA Received01/04/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/27/2016
03/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-