Model Number 722003
Device Problem Device Issue (2379)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse)¿s visual review of the system showed the footswitch was bent and not working properly.Philips conducted its investigation based on information received.Based on available information, the cause could be related to the use of anti-fatigue mats under the footswitches since the footswitch was designed to be used on a solid surface.After replacing the footswitch, the reported problem with the system was solved.Philips is working on adding a metal plate under the footswitch to prevent bending.(b)(4).
 
Event Description
Philips received a complaint from a customer that the footswitch was not working.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS HEALTHCARE
veenpluis 4-6
p.o. box 10.000
best 5680 DA
NL  5680 DA
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
dusty leppert
veenpluis 4-6
p.o. box 10.000
best 5680 -DA
NL   5680 DA
MDR Report Key5343742
MDR Text Key35351601
Report Number3003768277-2016-00003
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial
Report Date 12/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number722003
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2015
Initial Date FDA Received01/05/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-