(b)(4).Investigation results: visual evaluation of the returned device showed that there were multiple bends and kinks throughout the working length.Functional evaluation found the brush would not extend.Device was disassembled.Assessment found one strand of the pull wire had been broken at about 4 mm from the hypotube.The bend and kinks in the working length were most likely due to some tortuous anatomical or some operational aspect of the procedure which restricted the pull wire¿s ability to move freely within the catheter.Attempts to extend and retract the brush with pull wire movement restricted resulted in the pull wire strand breaking at the hypotube.Therefore, the most probably root cause for this event is determined to be operational context.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database confirmed that no other complaints exist for the specified lot.
|