TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801074 |
Device Problem
Out-Of-Box Failure (2311)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/14/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Evaluation is in progress, but not yet concluded.The voltage from the o2 sensor with 100% oxygen was 1.340 volts direct current (vdc), which is within specifications.With the blender setpoint at 21%, the central control monitor (ccm) reported 30.5%, which is out of specifications.The external o2 analyzer showed the actual o2 as 20.9%.The srt re-calibrated and received the same results.
|
|
Event Description
|
Upon receipt of the device, the service repair technician (srt) reported that the oxygen (o2) sensor failed the reading accuracy portion of the verification test for the electronic patient gas system (epgs).There was no patient involvement.
|
|
Manufacturer Narrative
|
Per evaluation, the oxygen (o2) sensor passed voltage but failed o2 reading accuracy (complaint condition was duplicated).The voltage reading was 1.49 volts which was within specification.With the blender setpoint of 21%, the central control monitor (ccm) reading was 26.8% and o2 analyzer reading was 20.8% (out of specification).With the blender setpoint of 30%, the ccm reading was 31.0% and o2 analyzer reading was 25.6% (out of specification).With the blender setpoint of 80%, the ccm and o2 analyzer readings were in acceptable range and met specification.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|