During review of the in-house programming history database, it is suspected that a faulted system diagnostic test occurred after (b)(6) 2015.On (b)(6) 2015 a system and normal mode diagnostics were performed.A final interrogation was performed confirming the correct settings.On the next recorded office visit, the settings were found to be at unintended parameters.The device was programmed to intended settings on (b)(6) 2014.No patient adverse events were reported.
|
Additional information was received from the physician assistant that patient's change in settings due to computer software error occurred on (b)(6) 2015.The physician assistant mentioned that she had trouble interrogating the patient's device on (b)(6) 2015.She was able to interrogate the settings initially but was unable to program.She performed troubleshooting by replacing the cables, securing cable connections and replacing the tablet/handheld devices.She then identified the issue to be due to wand battery depletion by performing the battery test.When the 9v wand battery was changed, the patient's generator was able to be programmed.The change in the patient's device settings were corrected the same visit.Product information was requested but were not provided.
|