• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 2088TC/52
Device Problems Structural Problem (2506); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/17/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the asymptomatic patient presented in clinic for routine post operation check.Upon interrogation, the right ventricular lead exhibited sensing anomaly.The physician believed the lead had dislodged.During lead revision, fluoroscopy revealed the lead had not moved.Fluid in the inner lumen of the lead was noted and was thought the lead had abraded.The lead was explanted and replaced.The patient was in good condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key5355671
MDR Text Key35493653
Report Number2017865-2016-00164
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Model Number2088TC/52
Device Lot NumberA000014943
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/17/2015
Initial Date FDA Received01/08/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/19/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age78 YR
-
-