U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIS INTERNATIONAL ICHEM VELOCITY; AUTOMATED URINE CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIS INTERNATIONAL ICHEM VELOCITY; AUTOMATED URINE CHEMISTRY ANALYZER Back to Search Results
Catalog Number 700-7177-001
Device Problem Fluid/Blood Leak (1250)
Patient Problem Not Applicable (3189)
Event Date 12/14/2015
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) evaluated the instrument.The fse found that the tubing at the color gravity module fitting was damaged.The fse replaced the tubing, which repaired the leak.The repairs were verified by the fse.(b)(4).
 
Event Description
The customer reported a leak from the ichem velocity instrument.The volume of the leak was not specified but was contained within the instrument.The customer was wearing personal protective equipment (ppe) consisting of gloves, goggles and a laboratory coat at the time of the occurrence and there was no report of injury or direct exposure to the leak.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY
Type of Device
AUTOMATED URINE CHEMISTRY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-s08
miami, FL 33196
3053802031
MDR Report Key5361915
MDR Text Key35733658
Report Number2023446-2016-00015
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2015
Initial Date FDA Received01/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-