• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS HEALTHCARE ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS HEALTHCARE ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number 722003
Device Problems Loss of Power (1475); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/01/2015
Event Type  Injury  
Manufacturer Narrative
Investigation is ongoing.
 
Event Description
The customer reported that the system went down while there was a patient under anesthesia on the table.
 
Manufacturer Narrative
Philips investigated this complaint and came to the following conclusion: the hemodynamic personal computer was defect, as a result of a power blackout test performed by the customer the previous day.This system was connected to the flexvision of the allura system, resulting in no visible hemodynamic data on the flexvision monitor.The hemodynamic system itself also carry its own monitor for hemodynamic information.This pc was replaced by the field service engineer.The live image on the flexvision was still available.No patient harm was reported due to this issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS HEALTHCARE
veenpluis 4-6
p.o. box 10.000
best 5680 DA
NL  5680 DA
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
dusty leppert
veenpluis 4-6
p.o. box 10.000
best 5680 -DA
NL   5680 DA
MDR Report Key5369636
MDR Text Key36057637
Report Number3003768277-2016-00007
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number722003
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/18/2015
Initial Date FDA Received01/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/18/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-