The subject device was returned to olympus (olympus (b)(4) sales & service co.) for evaluation.Olympus confirmed the subject device worked properly.Olympus did not determine the cause of this phenomenon at that time because the phenomenon was not reproduced under the investigation.Olympus also checked the device history record of the subject device, and there was no irregularity found.There were no further details provided.If significant additional information is received, this report will be supplemented.
|