• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENT/TRIAL Back to Search Results
Catalog Number 221750041
Device Problem Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/06/2016
Event Type  malfunction  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.Udi: (b)(4) unknown.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Surgeon wanted to reposition cup after initially inserting acetabular cup.The hole at end of handle was damaged and the slap hammer extractor could not be screwed into the hole to extract acetabular cup.
 
Manufacturer Narrative
Conclusion and justification status for mdr: the devices associated with this report were not returned.Based on the failure noted and previous evaluations, including evaluations by the depuy metallurgy team, the root cause is attributed to being repeatedly loaded in rotating bending fatigue beyond the material limit.No evidence of manufacturing or material defects has been observed that could contribute to the failure of these components.Based on the performed investigation, corrective action is not indicated at this time.Continue to monitor via sep-419.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5377221
MDR Text Key36631846
Report Number1818910-2016-11169
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PP070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number221750041
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2016
Initial Date FDA Received01/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/24/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-