The investigation determined that a non-reproducible, higher than expected vitros k+ result was obtained from a single patient sample using vitros k+ reagent processed on a vitros 5600 integrated system.The assignable cause of the event is unknown; however, a pre-analytical sample processing or a sample quality issue cannot be ruled as a contributing factor.The investigation was able to rule out a vitros k+ reagent issue or vitros instrument issue as possible contributing factors.The lower than expected vitros k result was reported outside of the laboratory, however, there was no allegation of patient harm as a result of this event.
|