Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the manifold was leaking where the plastic meets metal, and the sieve caps were leaking causing the sieve beds to be saturated, causing the 3/4 alarm, which confirmed the original complaint issue.However, there was no indication of a failure of the audible alarm to operate.Additionally, the battery had a dead cell causing it not to charge, verifying the complaint of the unit not turning on unless the charger was plugged in.
|