MAUDE Adverse Event Report: LIEBEL FLARSHEIM HUT EXT DR FINAL ASSY-STANDARD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIEBEL FLARSHEIM HUT EXT DR FINAL ASSY-STANDARD Back to Search Results
Model Number HUT EXT DR FINAL ASSY-STANDARD
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/16/2015
Event Type  malfunction  
Manufacturer Narrative
Biomed called service reporting that the collimator blades were stuck and he could not get them clear.Biomed asked for a field service engineer (fse) to come onsite.On (b)(6) 2015 fse went on site, he confirmed blade issue, replaced the collimator, and checked the unit for proper operation before returning unit to the customer for full service.
 
Event Description
Customer report that during an unknown procedure the collimator failed disabling fluoro.Collimator override did not work and the blades were stuck in the closed position.Staff was able to complete the procedure with the endoscope.No reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUT EXT DR FINAL ASSY-STANDARD
Type of Device
HUT EXT DR FINAL ASSY-STANDARD
Manufacturer (Section D)
LIEBEL FLARSHEIM
2111 east galbraith road
cincinnati OH 45237
Manufacturer Contact
david benson
2111 east galbraith road
cincinnati, OH 45237
5139485719
MDR Report Key5398128
MDR Text Key37231433
Report Number1518293-2015-00127
Device Sequence Number1
Product Code IXR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Biomedical Engineer
Device Model NumberHUT EXT DR FINAL ASSY-STANDARD
Device Catalogue Number404008
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/16/2015
Initial Date FDA Received01/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-