• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-RAD MEDICAL DIAGNOSTICS GMBH REAGENT RED BLOOD CELLS BIOTESTCELL 3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-RAD MEDICAL DIAGNOSTICS GMBH REAGENT RED BLOOD CELLS BIOTESTCELL 3 Back to Search Results
Catalog Number 816 085 100
Device Problems Device Expiration Issue (1216); False Negative Result (1225); Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
This is our combined initial and final report on this incident.
 
Event Description
The customer reported false negative test results with biotest cell 3 when used on tango optimo.The customer reported that the antibodies anti-e, anti-jk(b) and anti-le(a) were missed.The customer did not provide a date of event.The customer did neither return the patient samples that had caused false negative test results nor the supposedly defective product.At the time the customer filed his complaint, the supposedly defective product was already expired.Therefore our quality control laboratory did not test their retained biotest cell 3 sample.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.The affected tango optimo was inspected by a field service engineer and was found to be working within its specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAGENT RED BLOOD CELLS BIOTESTCELL 3
Type of Device
BIOTESTCELL 3
Manufacturer (Section D)
BIO-RAD MEDICAL DIAGNOSTICS GMBH
industriestrasse 1
dreieich, hessen 63303
GM  63303
Manufacturer (Section G)
BIO-RAD MEDICAL DIAGNOSTICS GMBH
industriestrasse 1
dreieich, hessen 63303
GM   63303
Manufacturer Contact
martina benkert
industriestrasse 1
dreieich, hessen 63303
GM   63303
103 3130 5
MDR Report Key5406288
MDR Text Key37790419
Report Number9610824-2016-00002
Device Sequence Number1
Product Code KSZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2015
Device Catalogue Number816 085 100
Device Lot Number854102-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2016
Initial Date FDA Received02/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-