• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE 2 DOCKING STATION (120V); APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE 2 DOCKING STATION (120V); APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Catalog Number 0702014000
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/07/2016
Event Type  malfunction  
Manufacturer Narrative
After the field service technician finished the analysis and safety testing, the device was returned to service at the customer facility.
 
Event Description
It was reported that during docking at the user facility, the waste hose of the docker came detached from the waste receptacle of the user facility, resulting in surgical waste being discharged into the room.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPTUNE 2 DOCKING STATION (120V)
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key5406607
MDR Text Key37427669
Report Number0001811755-2016-00125
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0702014000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/07/2016
Initial Date FDA Received02/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/22/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-