• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 CWS 400 CLOSED WOUND SUCTION KIT, PVC DRAIN; 3 SPRING EVACUATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 CWS 400 CLOSED WOUND SUCTION KIT, PVC DRAIN; 3 SPRING EVACUATOR Back to Search Results
Catalog Number 0043610
Device Problem Obstruction of Flow (2423)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Device not returned.
 
Event Description
It was reported that there was clotting in the end of the tubing that connects to the suction port.There was no patient impact.The device was replaced and worked without issue.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown therefore the device history record could not be reviewed.The instructions for use state the following: "precautions: if the drain is occluded, irrigation and/or aspiration of the drain may be required.Warnings: in the event of occlusion of the drain, all wound drainage ceases.Careful attention to the drain will minimize the probability of this problem.If occlusion does occur, the drain can be aspirated by connecting auxiliary suction to the reservoir outlet or temporarily disconnecting the drain from the evacuator and applying auxiliary suction directly to the drain." (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CWS 400 CLOSED WOUND SUCTION KIT, PVC DRAIN
Type of Device
3 SPRING EVACUATOR
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key5410391
MDR Text Key37521650
Report Number1018233-2016-00086
Device Sequence Number1
Product Code GCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 03/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0043610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/13/2016
Initial Date FDA Received02/04/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received02/25/2016
03/21/2016
04/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-