MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA PRESSURE WIRE AERIS G8; TRANSDUCER, PRESSURE,CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA PRESSURE WIRE AERIS G8; TRANSDUCER, PRESSURE,CATHETER TIP Back to Search Results
Model Number C12058
Device Problems Fracture (1260); Material Fragmentation (1261); Kinked (1339)
Patient Problem No Information (3190)
Event Type  Injury  
Event Description
Attempted ffr measurement with a pressurewire aeris in the lad via the right groin was unsuccessful.Due to tortuous anatomy, the wire became kinked in the right iliac artery.During removal of the pressurewire aeris, the wire fractured and a fragment remained in the artery.Ffr was then attempted via the left iliac artery but was unsuccessful.During the second attempt at ffr, the remaining fragment in the right iliac artery was visualized by angiography.After abortion of the ffr procedure, the fragment was snared, but the snared piece fragmented again.A small portion of the original fragment was removed with the snare.The remaining fragment was removed via surgical cut down.
 
Manufacturer Narrative
(b)(4).Product evaluation: the results of the investigation are inconclusive since the device was not returned for analysis.Our investigation was limited to the review of the device history record, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with sjm specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.The pressurewire instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire instructions for use (ifu) states that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE WIRE AERIS G8
Type of Device
TRANSDUCER, PRESSURE,CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key5412677
MDR Text Key37539671
Report Number3008452825-2016-00010
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131452
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2017
Device Model NumberC12058
Device Catalogue NumberC12058
Device Lot Number5202229
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2016
Initial Date FDA Received02/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-