• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T180
Device Problems Over-Sensing (1438); Pacing Problem (1439)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/31/2005
Event Type  malfunction  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Laboratory analysis did not confirm the reported clinical observations.An initial report was previously submitted to fda on 02/01/2006; however due to emdr submission issue related to the supplemental report, this is being filed again as an initial report.(b)(4).
 
Event Description
Additional information was received that this icd was later explanted and replaced for reported normal battery depletion.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5413676
MDR Text Key37571344
Report Number2124215-2016-02328
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/31/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/25/2006
Device Model NumberT180
Other Device ID NumberVITALITY DR HE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 12/31/2005
Initial Date FDA Received02/05/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
T180
Patient Age67 YR
-
-