Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Laboratory analysis did not confirm the reported clinical observations.An initial report was previously submitted to fda on 02/01/2006; however due to emdr submission issue related to the supplemental report, this is being filed again as an initial report.(b)(4).
|