Device evaluation summary device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (exposed wires) has been confirmed.As received, the electrode belt failed incoming testing.Upon investigation, the cable connecting the distribution node (dn) to rear therapy pad (te) was pulled from the strain relief.The cause of the test failure and exposed wires was the pulled dn to rear te cable.The root cause of the pulled dn to rear te cable was excessive force.No adverse event resulted from the defective electrode belt.
|