CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A - UPGRADE
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/04/2014
Event Type  malfunction  
Manufacturer Narrative
This medwatch report was identified as a late submission during a two year retrospective review of complaints and mdr¿s following receipt of an untitled letter issued by the fda.The carefusion fse went onsite and installed and new patient circuit.Performed a circuit calibration and a performance checkout.Paw=22.5, deltap=58 w/o humidifier.Driver/bellows clamps are very tight and 2 are missing handles - ordered 4 new ones and will return.Back onsite (b)(4) 2014.Replaced 4 driver clamps.Performed a circuit calibration and a performance checkout.All is ok now.
 
Event Description
The following description of the event was documented by a carefusion tech support specialist in a response to a phone conversation with a user facility representative on (b)(6) 2014.The customer called to report that this vent "stopped twice while on a patient." the first time, they were able to restart the driver.The second time, they were not able to re-pressurize and restart the driver.They were able to switch the patient to another 3100a without incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
marshall thompson
7607787307
MDR Report Key5415189
MDR Text Key38209335
Report Number2021710-2016-02899
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A - UPGRADE
Device Catalogue Number76890
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/04/2014
Initial Date FDA Received02/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-