Model Number 3100B
Device Problems Sticking (1597); Pressure Problem (3012)
Patient Problem No Patient Involvement (2645)
Event Date 02/24/2014
Event Type  malfunction  
Manufacturer Narrative
A carefusion fse went onsite and installed a pr3 and the system pressurized correctly.The fse also installed a needle valve assembly.All is ok now.As of this date the parts have not been received for evaluation.When it has been received and evaluated a supplement will be filed.
 
Event Description
The following description of the event was documented by a carefusion tech support specialist in a response to a phone conversation with a user facility representative on (b)(6) 2014 the customer called to report that they can't get any pressure out of this vent.They also report that the adjust knob doesn't seem to move like it should.The customer also stated there was no patient involvement in this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
marshall thompson
7607787307
MDR Report Key5415312
MDR Text Key38158345
Report Number2021710-2016-02924
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/24/2014
Initial Date FDA Received02/06/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2004
Type of Device Usage Reuse
Patient Sequence Number1
-
-