Model Number 3100B
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 02/26/2014
Event Type  malfunction  
Manufacturer Narrative
This medwatch report was identified as a late submission during a two year retrospective review of complaints and mdr¿s following receipt of an untitled letter issued by the fda.The unit was received at carefusion for evaluation.The issue was isolated to the thermistor wire being disconnected from the choke board.
 
Event Description
The following description of the event was received from carefusion fse on (b)(6) 2014.The carefusion fse was testing the driver thermistor voltage and noticed that when he first measured the voltage, it was 6v.As he increased the settings and disconnected the cooling gas, the voltage went up to 9v and is now stuck there.It does not come down as the driver warms up.This was found during testing so there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
marshall thompson
7607787307
MDR Report Key5415314
MDR Text Key38159766
Report Number2021710-2016-02926
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number773967-RNT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2014
Initial Date FDA Received02/06/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-