• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A - UPGRADE
Device Problem Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/26/2014
Event Type  malfunction  
Manufacturer Narrative
This medwatch report was identified as a late submission during a two year retrospective review of complaints and mdr¿s following receipt of an untitled letter issued by the fda.The carefusion fse went onsite and found no gas flowing from the pr5 cooling gas regulator.Lack of cooling gas caused repeated driver overheating.Due to potential damage to driver he the replaced driver and installed 6k pm kit which includes the pr5 regulator.I performed 6k pm, patient circuit calibration @ 40 cmh2o, and performance test @ 20.1 cmh2o paw / 64 cmh2o amplitude.As of this date the part has not been received for evaluation.When it has been received and evaluated a supplement will be filed.
 
Event Description
The following description of the event was documented by a carefusion tech support specialist in a response to a phone conversation with a user facility representative on (b)(6) 2014.The customer called to report that this unit is intermittently starting and stopping on its own.The unit seems to pressurize fine, the performance check seems to come in fine, but now the oscillator will stop on its own and alarms.The amps drop to zero and the map seems to stay stable.Then the unit will start up again on its own without her depressing the start/stop button.The hz drops from 15 to 11 hz as well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
marshall thompson
7607787307
MDR Report Key5415626
MDR Text Key38228326
Report Number2021710-2016-02942
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A - UPGRADE
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2014
Initial Date FDA Received02/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/1997
Type of Device Usage Reuse
Patient Sequence Number1
-
-