• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/23/2013
Event Type  Injury  
Manufacturer Narrative
This initial report was identified as a late submission during a two year retrospective review of complaints and mdr¿s following receipt of an untitled letter issued by the fda.(b)(4).The device/components documented as returned, failure analysis evaluation is not available.
 
Event Description
The customer reported the c30 on the unit was burnt and requested a new alarm board.The unit seems to run fine and alarms correctly but he needs a replacement because it's damaged now.The unit was running on a patient and they smelled something electrical burning.He states they swapped the unit out with another 3100a and there was no patient compromise.A replacement alarm board was issued and a return issued to sent defective alarm board for evaluation.
 
Manufacturer Narrative
Failure analysis (fa) lab received an alarm board assy (3100a).Performed alarm function checkout.The fa lab performed an investigation and the system worked as normal for 2 hours and no problems were found.The fa lab was able to duplicate the reported issue of c30 was burned on alarm board.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
7149227830
MDR Report Key5418216
MDR Text Key38249709
Report Number2021710-2016-02984
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 09/23/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2013
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/23/2013
Initial Date FDA Received02/08/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received08/19/2016
12/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-