Model Number ETH2B5LT |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/16/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not yet been returned to the manufacturer at the time of this report.A supplemental form will be sent once the evaluation is completed if the device is returned.The device history report was reviewed and no discrepancies were found.
|
|
Event Description
|
It was reported that when used during a laparoscopic appendectomy, it was noticed that the device had a piece of something black in it.The device was immediately removed from the patient and a new device was used.It was reported that there was no patient injury or consequences associated with this.
|
|
Manufacturer Narrative
|
The device was returned to the manufacturer and examined.Contaminants were found on the exterior of the device.No obvious debris was found within the inside of the cannulated obturator.The device was examined while being backlit and a small dot near its tip was found when viewed extremely close.The exterior contaminants were cleaned away, and the device was sonicated within a graduated cylinder.No contaminants were located within the sonicated water or were found to be expelled from the interior of the obturator.Upon further examination the dot was still present.The obturator had its tip removed for a close examination of the object.Several air bubbles were noted in the tip of the device once removed.No debris was found.The object in the account's pictures is believed to be an oem deformation which, when back lit cast the shadow during use in the field.No debris was found.Unable to confirm account complaint.
|
|
Search Alerts/Recalls
|