• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERILMED, INC.; LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERILMED, INC.; LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED Back to Search Results
Model Number ETH2B5LT
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/16/2016
Event Type  malfunction  
Manufacturer Narrative
The device has not yet been returned to the manufacturer at the time of this report.A supplemental form will be sent once the evaluation is completed if the device is returned.The device history report was reviewed and no discrepancies were found.
 
Event Description
It was reported that when used during a laparoscopic appendectomy, it was noticed that the device had a piece of something black in it.The device was immediately removed from the patient and a new device was used.It was reported that there was no patient injury or consequences associated with this.
 
Manufacturer Narrative
The device was returned to the manufacturer and examined.Contaminants were found on the exterior of the device.No obvious debris was found within the inside of the cannulated obturator.The device was examined while being backlit and a small dot near its tip was found when viewed extremely close.The exterior contaminants were cleaned away, and the device was sonicated within a graduated cylinder.No contaminants were located within the sonicated water or were found to be expelled from the interior of the obturator.Upon further examination the dot was still present.The obturator had its tip removed for a close examination of the object.Several air bubbles were noted in the tip of the device once removed.No debris was found.The object in the account's pictures is believed to be an oem deformation which, when back lit cast the shadow during use in the field.No debris was found.Unable to confirm account complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED
Manufacturer (Section D)
STERILMED, INC.
11400 73rd ave n
maple grove MN
Manufacturer Contact
jason anderson
5010 cheshire parkway
suite 2
plymouth, MN 55446
7634888348
MDR Report Key5419422
MDR Text Key38455313
Report Number2134070-2016-00009
Device Sequence Number1
Product Code NLM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2016
Device Model NumberETH2B5LT
Device Catalogue NumberETH2B5LT
Device Lot Number1895306
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2016
Initial Date FDA Received02/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-