• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON RECHARGEABLE IPG, 16-CHANNEL; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON RECHARGEABLE IPG, 16-CHANNEL; SCS IPG Back to Search Results
Model Number 3716
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 01/18/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).This ipg serial number was included in a field correction.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
Device 1 of 2: reference mfr.Report: 1627487-2016-00622.The patient reported experiencing heating at the ipg site while charging the ipg.A low energy charging system was sent to the patient which resolved the issue.On (b)(6) 2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON RECHARGEABLE IPG, 16-CHANNEL
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5420138
MDR Text Key37803182
Report Number1627487-2016-00623
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 01/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2014
Device Model Number3716
Device Lot Number3595478
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/18/2016
Initial Date FDA Received02/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/19/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Treatment
MODEL 3219, SCS LEAD
Patient Outcome(s) Other;
Patient Age61 YR
-
-