During an internal review of programming and diagnostic data, it was found that the vns patient's device was tested on (b)(6) 2014 and system mode diagnostic results revealed high impedance (dc/dc code 7).The device was then switched off.On (b)(6) 2015, the device was turned on and new tests were run: system diagnostics returned again high impedance with dc/dc code 7.Then the device was again switched off.Further information was received from the physician, indicating that the patient's device was definitely switched off on (b)(6) 2015 because it did not seem to have provided much if any clinical benefit since it was implanted.It was reported that the device was not explanted since, as the patient is not keen to have it explanted.No known surgical interventions have occurred to date.
|