MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problem Impedance Problem (2950)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/11/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The healthcare provider (hcp) via the representative reported impedance out of range upon connection.It was unknown what led to the event, but multiple attempts with reconnecting were tried.To resolve the issue, a new device was used.It was noted the device was never implanted to the pocket.
 
Manufacturer Narrative
(b)(4).Analysis of the ins ((b)(4)) found no anomaly.The ins passed all functional testing including a test of the impedance measurement function.The impedances were 510 ohm loads connected across all three electrodes.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key5428100
MDR Text Key38529005
Report Number3004209178-2016-02591
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2017
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2016
Initial Date FDA Received02/10/2016
Supplement Dates Manufacturer ReceivedNot provided
02/15/2016
Supplement Dates FDA Received03/09/2016
09/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00049 YR
-
-