ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Patient Involvement (2645)
Event Date 01/20/2016
Event Type  malfunction  
Manufacturer Narrative
The system was examined and the reported event was confirmed (the system would not respond).The host module was replaced to address the issue.The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on (b)(6) 2014.A review of the manufacturing records did not reveal any related non-conformity during manufacturing for this system.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the non-conforming host.However, how and when the host module became nonconforming cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Event Description
An ophthalmic surgeon reported that prior to a vitrectomy procedure the system locked up.The surgery was initiated and completed using an alternate system.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key5429940
MDR Text Key38638538
Report Number2028159-2016-00604
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number3.00.61
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2016
Initial Date FDA Received02/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-