• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H802228240022
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 11/02/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by manufacturer: examination of the retuned complaint device revealed that the unit was received in a sealed pouch, it was visually inspected through the pouch and a foreign matter was noted inside the package.The foreign matter was measured and did not meet the measure test specification since the results were over of 1mm².The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause has been determined to be manufacturing related.(b)(4).
 
Event Description
Reportable based on device analysis completed on (b)(4) 2016.It was reported that foreign material was noted during preparation.This rotawire and wireclip torquer was selected; however, during unpacking they found "gray like dust" inside the exterior package.The procedure was completed with a different device.No patient complications were reported and that patient's status is good.However, device analysis revealed a foreign matter inside the sealed pouch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE¿ AND WIRECLIP¿ TORQUER
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5430258
MDR Text Key38002073
Report Number2134265-2016-00463
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 01/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/23/2017
Device Model NumberH802228240022
Device Catalogue Number22824-002
Device Lot Number17820388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/13/2016
Initial Date FDA Received02/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial