BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY
|
Back to Search Results |
|
Model Number H802228240022 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/02/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Device evaluated by manufacturer: examination of the retuned complaint device revealed that the unit was received in a sealed pouch, it was visually inspected through the pouch and a foreign matter was noted inside the package.The foreign matter was measured and did not meet the measure test specification since the results were over of 1mm².The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause has been determined to be manufacturing related.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on (b)(4) 2016.It was reported that foreign material was noted during preparation.This rotawire and wireclip torquer was selected; however, during unpacking they found "gray like dust" inside the exterior package.The procedure was completed with a different device.No patient complications were reported and that patient's status is good.However, device analysis revealed a foreign matter inside the sealed pouch.
|
|