It was reported during prophylactic generator replacement, the surgeon found the lead wiring out of the tubing.When the surgeon went to remove the generator from the pocket the lead broke in half.It was reported that device diagnostics prior to generator replacement were within normal limits.The explanted generator and lead were received for analysis.Analysis of the generator was completed on 02/08/2016.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery shows an ifi=no condition.There were no performance or any other type of adverse conditions found with the pulse generator.Analysis of the lead was completed on 02/02/2016.Note that the electrodes were not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis of the returned lead portions the quadfilar coils did not appeared to be broken and abraded openings were not observed on the outer or inner silicone tubes.Slice marks were observed in various areas on the outer silicone tubing.The slice marks found on the outer silicone tubing and the cut ends that were made during the explanted process, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids found inside the outer silicone tubing.What appeared to be white deposits were observed in various locations.The condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The setscrew marks found on the connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, with no discontinuities identified.Based on the analysis results of the lead, it is likely that the damage to the lead occurred as a result of the surgery; however, attempts to obtain additional relevant information have been unsuccessful to date.
|