MAUDE Adverse Event Report: NIHON KOHDEN TOMIOKA CORPORATION AG-920RA; MULTI-GAS UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN TOMIOKA CORPORATION AG-920RA; MULTI-GAS UNIT Back to Search Results
Model Number AG-920RA
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/12/2016
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the multi-gas unit was giving low readings.The biomedical engineer had not done the calibration, or dryline receptacle.The biomedical engineer had tried new water traps.This issue was discovered as the device was being checked before starting a case.The device was returned to nihon kohden and evaluated, but the malfunction could not be duplicated.All gases fell within the service manuals specifications.Unit passes air and gas calibration.This unit was tested several times and never had an issue.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the multi-gas unit was giving low readings.The biomedical engineer had not done the calibration, or dryline receptacle.The biomedical engineer had tried new water traps.This issue was discovered as the device was being checked before starting a case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AG-920RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn; susan shadley
tomioka city, japan 370-2 314
JA  370-2314
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn; susan shadley
tomioka city, japan 370-2 314
JA   370-2314
Manufacturer Contact
susan shadley
1-31-4 nishiochiai, shinjuku-k
attn: susan shadley
tokyo, japan 161-8-560
JA   161-8560
2687133
MDR Report Key5432812
MDR Text Key38081618
Report Number8030229-2016-00054
Device Sequence Number1
Product Code CCK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/11/2016,01/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAG-920RA
Device Catalogue NumberAG-920RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/11/2016
Distributor Facility Aware Date01/12/2016
Device Age48 MO
Event Location Hospital
Date Report to Manufacturer02/11/2016
Initial Date Manufacturer Received 02/11/2016
Initial Date FDA Received02/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-