Biomed called service and stated that a physicist was just there to check the system and the physicist said the system was reading a fluoro max dose of about 13r/min and he needs to adjust it down to just below 10r/min.Tech service explained to biomed how to adjust the max dose.Biomed was to call back if needed.Tech service reports that the hospital's biomed(s) take care of this system and they only call for a field service engineer (fse) on major service calls.The last time an fse was at this site was on (b)(6) 2015 service ticket: (b)(4).The last time one of liebel flarsheim's (lf) field engineers would have checked the dose was on service ticket: (b)(4), (b)(6) 2013 when service upgraded the x-ray generator and replaced the x-ray tube.At liebel flarsheim's (lf) request, service spoke with the biomed who reported the initial value for e17 = 140 with the dose at 13r/min.He dropped e17 = 132 for a dose of 9.4 r/min.
|