One triple lumen presep catheter kit with all components including packaging was returned for evaluation.As received, one side of shelf box was opened, and tyvek sheet was also opened from the tray.An insect was found stuck on the inner surface of tyvek sheet.The insect seemed to be crushed to tyvek sheet and a stain was also observed on the other side of tyvek sheet, where the insect was found.The insect was approximately 3mm x 4mm in size.No visible damage tyvek sheet or tray was observed.Visual examinations were performed under microscope at 10x magnification.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of ¿insect was stuck to the inner surface of the tyvek sheet¿ was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.This model number is exclusively sold in the (b)(4) market; therefore, this is a regional recall only.
|