It was reported by a nurse at the user facility that approximately 9 months ago, while the sst forceps were being used in surgery, the tip of the forceps broke and fell into the incision.The broken portion of the forceps was retrieved and the surgery completed with no injury to the patient.
|
The broken sst forceps was received for evaluation on 01/13/2016.Visual inspection confirmed the customer's report.The lower cutter was broken off.Additionally, the actuator was worn, the etching was discolored, and the instrument would not function.Based on the condition and age of this device; this reported breakage event was the result of excessive usage and general wear and tear.The tip/cutter, tube and actuator were replaced, the handle was refurbished, and the device has been returned to the customer.This device was manufactured on 29-jul-2004.This device is over twelve (12) years old.Extended use and handling of the device can cause wear, damage and/or breakage to the forceps.A review of the service history shows there have been no previous repairs or maintenance performed on this instrument.A review of complaint history shows there have been no previous complaints reported on this device.The product insert that accompanies this device provides the following precautions: inspect instrument prior to use to ensure it is in good physical condition and functions properly.There should be no loose, broken or misaligned parts.Exercise care in the use of the device to minimize side or bending loads.Do not use excessive force on instruments to avoid damage or breakage during use.Avoid unintended contact with other surgical instruments during use to prevent damage or breakage.Inspect instrument after use to ensure it has not been damaged.
|