Model Number 4FC12 |
Device Problems
Difficult to Advance (2920); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
|
|
Event Description
|
It was reported that during a cryoablation procedure, when inserting the balloon catheter into the sheath there was a strong resistance.The advancing of the balloon was attempted again without resolve.The balloon catheter was then replaced with a new one.While inserting the balloon catheter into the sheath, it was confirmed that air got into the sheath.Air continuously entered so a new sheath was used.The procedure was able to be completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Event summary: investigation was unable to be completed due to no product nor patient data file return.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|