• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VELA COMP D
Device Problems Delamination (2904); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 01/22/2016
Event Type  malfunction  
Manufacturer Narrative
Carefusion complaint number (b)(4).In the event the device is received for evaluation or additional information is received a follow-up report will be submitted.Carefusion has yet to receive the sample for evaluation.(b)(4).
 
Event Description
The customer stated that this unit has a non-responsive, delaminating screen.There was no patient involvement at the time of the event.
 
Manufacturer Narrative
Results of investigation: the carefusion failure analysis lab evaluated the front panel and was able to reproduce the reported event of there being fluid ingress on the panel screen, and when powering on screen calibration was unable to be performed.This issue is being addressed through an internal investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELA VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch pkwy
yorba linda, CA 92887
7149227830
MDR Report Key5445698
MDR Text Key38701655
Report Number2021710-2016-03164
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K093094
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVELA COMP D
Device Catalogue Number16532-02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2016
Initial Date FDA Received02/18/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-