Model Number IRC5PO2V |
Device Problems
Device Alarm System (1012); Out-Of-Box Failure (2311)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/03/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product has not been returned for evaluation as of the date of this investigation.Rma was issued.If new information becomes available at a later date this complaint will be updated and/or a follow up be sent.
|
|
Event Description
|
Provider states out of box, the concentrator alarm will not function.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that no alarms are working due to the pcb, which confirmed the original complaint issue.However, the underlying cause could not be determined.Fields were updated accordingly.
|
|
Event Description
|
Provider states out of box, the concentrator alarm will not function.
|
|