COVIDIEN 980 VENTILATOR; CONTINUOUS VENTILATOR Back to Search Results
Model Number 980
Device Problem Poor Quality Image (1408)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The covidien service engineer (se) inspected the device and verified the reported malfunction.The se replaced the graphic user interface to resolve the issue.The se performed extended self-testing on the device and all tests passed.
 
Event Description
It was reported that the screen on a 980 ventilator was dim and almost unreadable.The ventilator was not in use on a patient at the time the malfunction occurred.
 
Manufacturer Narrative
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5454577
MDR Text Key39215191
Report Number8020893-2016-00413
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K131252
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number980
Device Catalogue Number980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/01/2016
Initial Date FDA Received02/23/2016
Supplement Dates Manufacturer Received02/01/2016
Supplement Dates FDA Received12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-