Model Number 10-80-00
Device Problem Failure to Pump (1502)
Patient Problem No Patient Involvement (2645)
Event Date 02/05/2016
Event Type  No Answer Provided  
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the s5 roller pump displayed a pump failure alarm during setup that could not be cleared.There was no patient involvement.A sorin group field service advised the perfusionist to move the arterial pump to a lower-prioirity position until the device could be investigated on site.The service representative was dispatched to the facility and was informed that the pump had been run in subsequent cases without any other issues.The service representative ran the pump without incident and opened it up to check the connections.No issues were found.A pump readout was performed at the pump was reassembled and released to the customer.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the s5 roller pump displayed a pump failure alarm during setup that could not be cleared.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada 80004
3034676461
MDR Report Key5459139
MDR Text Key38972799
Report Number9611109-2016-00109
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 02/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Device Catalogue NumberN/A
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/05/2016
Initial Date FDA Received02/24/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/22/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-