Model Number SC-2316-70 |
Device Problems
Bent (1059); Break (1069); Fracture (1260); High impedance (1291); Kinked (1339)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 01/07/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Additional suspect medical device components involved in the event: model#: sc-2316-50 serial#: (b)(4) description: infinion1x16 perc lead kit-50 cm model#: sc-3400-30 serial#: (b)(4) description: infinion splitter 2x8 kit (30 cm).
|
|
Event Description
|
A report was received that the patient has several contacts on both leads that were measuring high impedances.Database analysis on impedance measurements revealed eleven contacts with high values on port ab.The patient underwent a revision procedure wherein the leads and splitters were replaced.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
(b)(4).Device evaluation indicated that the complaint had been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that multiple cables were completely broken at the bent/kinked location of the lead.The bent/kinked location is 1 cm from the set screw mark of the clik anchor.There were no exposed cables at the clik site fracture locations.The broken cables resulted in the reported complaint of high impedance.Sc-2316-70, (b)(4): device evaluation indicated that the complaint had been confirmed.Impedance tests confirmed the complaint.High impedance readings were registered at contacts 6 and 11.External visual inspection of the lead found that the lead proximal end was fractured between contacts # 10 and 11.X-ray inspection revealed broken cables located between contacts number 10 and 11.No cables were exposed at the fracture site.The dhr review found no anomalies when the lead was manufactured.(b)(4): device evaluation indicated that the complaint was not verified.A test infinion lead was inserted into the splitter connector without any anomalies.Device exhibited normal device characteristics.
|
|
Event Description
|
A report was received that the patient has several contacts on both leads that were measuring high impedances.Database analysis on impedance measurements revealed eleven contacts with high values on port ab.The patient underwent a revision procedure wherein the leads and splitters were replaced.The patient was doing well postoperatively.
|
|
Search Alerts/Recalls
|