The associated liner was returned and evaluated.The visual inspection of the returned device shows damage not unexpected for trying several attempts to insert.A dimensional inspection was attempted; however damage/deformation at several features of the device would not allow for accurate measurement.At this time, we have no reason to suspect that the product failed to meet any product specifications.This investigation could not verify or identify any evidence of product contribution to the reported problem.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.
|